The reported complaint that the autopulse platform (serial #(b)(4)) does not power on was confirmed during functional testing and visual inspection.The root cause of the reported complaint was due to a damaged battery cable in the battery compartment.The physical damaged appeared to be the characteristics of harsh impact due to user mishandling.Further visual inspection of the returned autopulse platform revealed a cracked front enclosure, unrelated to the reported complaint.The observed physical damage appeared to be the characteristics of harsh impact due to user mishandling.The front enclosure needs to be replaced to address the issues.The data archive shows no discrepancy.During functional testing, the autopulse platform failed to power up, thus confirming the reported complaint.The root cause of the reported complaint was due to a damaged battery cable in the battery compartment.The battery cable needs to be replaced to remedy the reported power issue.Zoll is awaiting on customer's approval for service repair.Historical complaints were reviewed for service information related to the reported complaint, and there was no previous history complaint reported for autopulse platform with serial number (b)(4).
|
Customer reported that the autopulse platform (serial # (b)(4)) does not power on and the pins inside the battery compartment are bent.It is unknown where the problem occurred.However, patient use information was requested, but no additional information was provided.
|