Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, during the beginning of the procedure, the confidence system leaked at the connection of the pump to the cement container.Another set was used to complete the procedure.There was a surgical delay of ten (10) minutes.The procedure was completed successfully.There was no patient confidence.This report involves one (1) confidence spinal cement system confidence kit spinal cement system.This is report 1 of 1 for (b)(4).
|
Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H4, h6: a manufacturing record evaluation was performed for the finished device product code: 283913000 lot: 317893 it was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: july 20, 2021 cement number: product code:183901001 / batch:9781015 the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned photo.Visual analysis of the photo provided did not find evidence of water leakage for the reported confidence kit, no needles.No indication of a significant product defect was found.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed for confidence kit, no needles.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|