The reported event was confirmed cause manufacturing related.One sample exhibited the reported failure.The device had not met specifications.The reported failure is considered out of specification as the reported failure was reproduced.The product was not used for patient treatment or diagnosis.The product caused the reported failure.A potential root cause for this failure could be inadequate pressure on the machine to punch.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The investigation was concluded, and no additional action is required at this time.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿warning: do not use ointments or lubricants having a petrolatum base.They will damage silicone.Visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.For urological use only.Do not use if package is damaged.Bard, bardex, bardia, biocath and lubricath are trademarks and/or registered trademarks of c.R.Bard, inc.Single use only.Do not resterilize consult instructions for use.Manufacturer: to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.¿ h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the actual/suspected device was inspected.
|