• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12°; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12°; HF-RESECTION ELECTRODES Back to Search Results
Model Number WA22302D
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/10/2021
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for evaluation/investigation since it was reportedly discarded by the user facility.Therefore, the cause of the reported event could not be conclusively determined.However this is a known error pattern which was most likely caused by used related wear and tear.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf resection electrode without showing any abnormalities.The case will be closed on olympus side with no further actions.However, the reported phenomenon will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Event Description
Olympus was informed that during a therapeutic resection of cervical submucosal myoma procedure the loop wire of the resection electrode broke off and fell into the patient's body cavity.The broken fragments could be retrieved successfully.The intended procedure was successfully completed with a similar device without significant delay and there was no report about an adverse event or patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12°
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov 75124
EZ   75124
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662
MDR Report Key13223412
MDR Text Key285046276
Report Number9610773-2022-00049
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761051665
UDI-Public14042761051665
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K120418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 01/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA22302D
Device Catalogue NumberWA22302D
Device Lot Number1000078192
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age31 YR
Patient SexFemale
-
-