Model Number G247
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Device Overstimulation of Tissue (1991)
Event Date 12/04/2021
Event Type  Injury  
Event Description
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) heard beeping from the device and noted a thumping feeling.Upon interrogation, the device exhibited error codes and was found to be operating in safety mode.The device was subsequently explanted and replaced.No additional adverse patient effects were reported.At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode and that both brady and tachy therapy remained available.Review of device memory identified an error.The error resulted in software resets performed in an attempt to correct an identified memory inconsistency.The corrupted memory was corrected in the laboratory and the device reverted to normal operation.The cause of the memory inconsistency was not able to be determined through laboratory testing but was likely the result of exposure to radiation, either therapeutic or environmental.
 
Event Description
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) heard beeping from the device and noted a thumping feeling.Upon interrogation, the device exhibited error codes and was found to be operating in safety mode.The device was subsequently explanted and replaced.No additional adverse patient effects were reported.At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13228114
MDR Text Key283622280
Report Number2124215-2022-00102
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589287
UDI-Public00802526589287
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/05/2023
Device Model NumberG247
Device Catalogue NumberG247
Device Lot Number260052
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age66 YR
-
-