Model Number CD3371-40C
Device Problem Product Quality Problem (1506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2021
Event Type  malfunction  
Event Description
It was reported that during a procedure prior to implant, markings were visually observed on the outside of the device.The device was not used and was replaced to resolve the event.The patient was in stable condition.
 
Manufacturer Narrative
The reported field event of can discoloration was confirmed by inspecting the device when received for analysis.The discoloration marks were analyzed and identified to be the parylene coating that was not entirely removed from the can surface during manufacturing.The anomaly is cosmetic in nature.The device functionality was tested in the lab and was revealed to be normal.Telemetry, pacing, sensing, impedance, high-voltage (hv) output, hv shock, and the patient notifier were all tested on the bench and no anomalies were revealed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13232149
MDR Text Key283686344
Report Number2017865-2022-00952
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Model NumberCD3371-40C
Device Lot NumberA000114554
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-