MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. OPSITE FLEXIFIX GENTLE 10CMX5M; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. OPSITE FLEXIFIX GENTLE 10CMX5M; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 66801197
Device Problem Loss of or Failure to Bond (1068)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that, during treatment with opsite flexifix gentle 10cm x 5m, when the carrier was removed, much of the silicone adhesive removed with the carrier and did not remain on the film, so it could not be used.This happened before use in patient.It is unknown how the treatment was performed.No harm to the patient or further complications reported.
 
Manufacturer Narrative
H3, h6: the device was returned for evaluation, confirming the reported event.Visual inspection noted no obvious defects, functional evaluation confirmed silicone remained on the carrier reducing adherence.The root cause has been deemed a component failure, the instructions for use, detail the wound contact layer within these dressing can be affected by storage temperature fluctuations.A documentation investigation has been conducted, confirming that no manufacturing problems have been observed.Complaint history shows previous events of this nature, with corrective actions assigned.The complained product was released after the actions being initiated.The instructions for use and risk files, mitigate the reported issue with no updates required.Smith and nephew can confirm the device was released according to specifications and continue to monitor for adverse trends relating to this product range.This investigation is now complete with no additional corrective actions deemed necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPSITE FLEXIFIX GENTLE 10CMX5M
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13236938
MDR Text Key284294064
Report Number8043484-2022-00027
Device Sequence Number1
Product Code KGX
UDI-Device Identifier05000223480400
UDI-Public5000223480400
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/01/2021
Device Catalogue Number66801197
Device Lot Number1368851950
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2022
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/28/2021
Initial Date FDA Received01/12/2022
Supplement Dates Manufacturer Received01/26/2022
Supplement Dates FDA Received01/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-