It was reported that an unknown patient underwent atrial fibrillation (afib) ablation procedure with a unk_smartablate pump tubing.There was a tubing problem during the procedure.It was reported that the smarttouch sf catheter was not irrigating properly.The catheter was flushed, without resolution.The caller reported that when coming on ablation, it was noticed that the temperature would immediately rise, and exceed the maximum temperature setting, cause ablation to cut off.The smarttouch sf catheter was removed from the body, and flushed, but the catheter was still not irrigating properly.The smarttouch sf catheter was replaced, and the issue still persisted.The reporter then checked the irrigation line, and it was noticed that the irrigation line or tubing had been "pulled up" into the smartablate pump clear plastic housing, causing the irrigation issue.The irrigation tubing was straightened out properly, and the issue had resolved.The procedure continued.No patient consequences were reported.Tubing problem during procedure is mdr-reportable.
|
On 26-mar-2022, the product investigation was completed as the complaint device was not returned.Since no device has been received for analysis, no product investigation can be performed, and the customer complaint cannot be confirmed.Manufacturer record evaluation cannot be conducted because the no lot number was provided by the customer.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|