• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LENSTEC BARBADOS INC. SOFTEC I; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LENSTEC BARBADOS INC. SOFTEC I; INTRAOCULAR LENS Back to Search Results
Model Number SOFTECI+20.0
Device Problem Device Handling Problem (3265)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 12/01/2021
Event Type  Injury  
Manufacturer Narrative
A full audit of all batch documentation relating to the production of the device was performed.The audit concluded that all procedures in manufacturing and packaging of the device had been conducted correctly.Batch reconciliation was 100.0%.The visual inspection performed on the device determined that the lens was damaged due to being improperly handled.
 
Event Description
Lenstec received an email stating " lens did not load properly in cartridge.Md attempted to insert the lens, but it was damaged.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFTEC I
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church,
BB 
Manufacturer (Section G)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church,
BB  
Manufacturer Contact
jimmy chacko
1765 commerce ave. n.
st. petersburg, FL 33716
7275712272
MDR Report Key13250940
MDR Text Key283758654
Report Number9613160-2022-00004
Device Sequence Number1
Product Code HQL
UDI-Device Identifier00844369029445
UDI-Public00844369029445
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P090022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberSOFTECI+20.0
Device Catalogue NumberSOFTEC I
Device Lot NumberSB12710
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2021
Initial Date Manufacturer Received 12/16/2021
Initial Date FDA Received01/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-