A us distributor contacted zoll to report that a patient passed away on (b)(6) 2021.It was reported that the patient was in the hospital at the time of passing.Review of the patient's download data revealed that prior to passing, the patient received two treatments from the lifevest, one appropriate and one inappropriate, followed by 6 non-lifevest defibrillations.At 22:21:05, the patient received the first treatment from the lifevest.The treatment was appropriate and was delivered while the patient's rhythm was vt at 220 bpm.The post-shock rhythm was sinus bradycardia at 40 bpm with hb and nsvt.At 22:25:41, the patient received the second treatment from the lifevest.The treatment was inappropriate and was delivered while the patient's rhythm was obscured by cpr and motion artifact.The post-shock rhythm was vt at 250 bpm with motion artifact.At 22:26:01, the first non-lifevest defibrillation was delivered.The patient's rhythm at the time of the treatment was vt at 260 bpm.The post-shock rhythm was obscured by cpr and motion artifact.The patient's rhythm then degraded to vf with cpr/motion artifact/electrode lead fall off.At 22:28:07, the patient received the second non-lifevest defibrillation.The patient's rhythm at time of treatment vf with cpr/motion artifact/electrode lead fall off.The post shock rhythm cpr/electrode lead fall off.The rhythm then transitioned to bradycardia at 50 bpm with cpr/electrode lead fall off.The rhythm then degraded further to vf with cpr/electrode lead fall off.At 22:31:02, the patient received the third non-lifevest defibrillation.The patient's rhythm at time of treatment vf with cpr/motion artifact/electrode lead fall off.The post shock rhythm was obscured by electrode lead fall off.The rhythm then degraded to vf with cpr/electrode lead fall off.At 22:23:15, the patient received the fourth non-lifevest defibrillation.The patient's rhythm at time of treatment was vf with cpr/motion artifact/electrode lead fall off.The patient's post shock rhythm was cpr/electrode lead fall off.The patient's rhythm degraded to vf with cpr/motion artifact/electrode lead fall off.At 22:35:42, the patient received the fifth non-lifevest defibrillation the patient's rhythm at time of treatment was vf with varying amplitudes, motion artifact, and electrode lead fall off.The patient's post shock rhythm was obscured by cpr/electrode lead fall off.The rhythm then degraded again to vf with motion artifact, varying amplitudes, and electrode lead fall off.At 22:37:43, the patient received the sixth non-lifevest defibrillation.The patient's rhythm at time of treatment was vf with motion artifact, varying amplitudes, and electrode lead fall off.The post shock rhythm was obscured by cpr/electrode lead fall off.The patient was last seen in an idioventricular rhythm from 20-40 bpm with cpr/motion artifact and electrode lead fall off until the device shutdown at 22:42:02 on (b)(6) 2021.
|