ST PAUL CADD; PUMP, INFUSION Back to Search Results
Model Number 21-2111-0300-01
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection did found the smiths tampered seal label intact, scratched and cracks on lcd lens screen.The customer stated problem was not confirmed.Unable to repeat, duplicate or determine the intermittent of the reported problem.Replaced the uso (upstream occlusion sensor) and dso sensor (downstream occlusion sensor) as a precaution measure.The root cause could not be determined.A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.
 
Event Description
It was reported the device had an intermittent infusion problem.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
receiving
minneapolis, MN 55442
MDR Report Key13261336
MDR Text Key284510381
Report Number3012307300-2022-00800
Device Sequence Number1
Product Code MEA
UDI-Device Identifier10610586038778
UDI-Public10610586038778
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130394
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number21-2111-0300-01
Device Catalogue Number21-2111-0300-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2021
Was the Report Sent to FDA? No
Date Manufacturer Received07/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-