It was reported that the right ventricular (rv) lead exhibited oversensing of intrinsic atrial beats resulting in multiple episodes with inappropriate anti-tachycardia pacing (atp) device therapy and inappropriate device shocks.Therapy was not exhausted.It was subsequently discovered that this cardiac resynchronization therapy defibrillator (crt-d) device had flipped multiple times within the pocket which pulled the leads and caused all the leads to dislodge from the heart.As a result, the rv lead, the right atrial (ra) lead and the left ventricular (lv) lead were all explanted and replaced, and this crt-d device was repositioned and remains in-service.No additional adverse patient effects were reported.
|