Catalog Number 2C2118K
Device Problem Failure to Prime (1492)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a small volume intermate would not flow.The device was filled with 0.9% sodium chloride.This issue was identified during priming.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured from september 07, 2021 - september 08, 2021.H10: the device was received for evaluation containing 80ml of fluid in the bladder.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.After the luer cap was removed and the slide clamp was disengaged from the tubing, evidence of continuous flow was visually observed at the distal luer.A functional flow rate test was performed and the flow rate was found to be within the product specification range.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERMATE
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong ave.
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13290840
MDR Text Key284187530
Report Number1416980-2022-00075
Device Sequence Number1
Product Code FRN
UDI-Device Identifier00085412487939
UDI-Public(01)00085412487939
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C2118K
Device Lot Number21J018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2022
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/24/2021
Initial Date FDA Received01/19/2022
Supplement Dates Manufacturer Received02/07/2022
Supplement Dates FDA Received02/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-