Model Number AIA-2000 ST
Device Problem Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2021
Event Type  malfunction  
Manufacturer Narrative
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by reviewing the error logs, but was not able to reproduced error.Fse verified liquid sense settings were in range, and the hybrid arm z-axis home sensor was functioning as expected.Fse proactively lubricated the guide rod and the spindle.Fse found misalignments at the hybrid arm and resolved the complaint by adjusting the hybrid arm alignments for the cup bottom positions of the reagent test cup and sample treatment cup lanes and also adjusted the reagent carousel small bottle bottom alignment for the hybrid arm to accurately calculate bottle volumes.Fse validated instrument by successfully performing quality control run without error and within acceptable range.No further action required by field service.The aia-2000 instrument is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were four (4) similar complaints identified during the searched period, which includes this event.The aia-2000 operator's manual under appendix 4: error messages states the following: [4253] hybrid arm z-axis home overrun cause : the home sensor activated improperly after movement of the hybrid arm z-axis.If retry fails, the measurement result will be flagged (mf flag).Solution : contact tosoh service center or local representatives.The most probable cause of the reported event is due to misaligned hybrid arm.
 
Event Description
A customer reported getting error message "4253 hybrid arm z-axis home overrun" on the aia-2000 instrument.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for intact parathyroid hormone (ipth).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
bernadette o connell
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368143
MDR Report Key13298767
MDR Text Key290530242
Report Number3004529019-2022-00011
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189284616
UDI-Public04560189284616
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000 ST
Device Catalogue Number022100
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/29/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-