H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned for evaluation.Based on the investigation performed in the laboratory it is confirmed that the user could only partially deploy the stent graft.Perforation of the inner layer of the outer sheath by one of the struts of the stent graft was identified as the main issue to the reported event which lead to partial deployment of the stent graft, and fracture of the outer sheath.The system was appropriately flushed and the anatomy of the patient was not calcified or tortuous in nature.Therefore, the investigation is confirmed for the reported issues.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.Regarding the preparation of the device the instructions for use states 'flush the stent graft lumen with sterile saline by using a small volume syringe.A) attach the syringe to the luer port at the back of the endovascular system (.).B) attach the syringe to the luer port on the y-adapter, (.) close the stopcock when flushing is complete and remove the syringe from the luer port'.Regarding the precautions prior and during deployment the instructions for use states 'prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure' and 'if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device¿.The instructions for use states regarding the accessories 'the use of an appropriately sized introducer sheath is recommended; prepare a stiff 0.035" guidewire per its instructions for use and advance the guidewire under fluoroscopy to the target location' the packaging pictograms indicate an introducer size of 9f and a 0.035" guidewire.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|