Model Number 72202536
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/05/2022
Event Type  malfunction  
Event Description
According to the reporter, at the end of the polypectomy procedure, the physician noticed some glitter like material in the area resected polyp.Physician successfully completed case and was pleased with outcomes. there was no patient injury.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted a solid yellow particulate matter.Functionally, the material was consistent with a build up of excess silicone lubricant and per the existing manufacturing work instructions associates shall verify that there was not a buildup of excess silicone on/in the inner blade.Blade discoloration was non-toxic, biocompatible, and safe for human use.If excess silicone was observed, associates shall remove the buildup using a wooden/plastic pick, then verify no nicks, dents, or scratches are present after removal.The device latched properly into the hand piece and functioned as indented.It was reported that the device produced shavings or fragments.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN MANSFIELD
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN MANSFIELD
15 hampshire street
mansfield MA 02048
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key13306905
MDR Text Key284116029
Report Number1282497-2022-00001
Device Sequence Number1
Product Code HIH
UDI-Device Identifier10884521744080
UDI-Public10884521744080
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130215
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72202536
Device Catalogue Number72202536
Device Lot Number5197650
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2022
Initial Date Manufacturer Received 01/05/2022
Initial Date FDA Received01/20/2022
Supplement Dates Manufacturer Received02/16/2022
Supplement Dates FDA Received02/23/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age34 YR
Patient SexFemale
-
-