The catalog number identified has not been cleared in the us but is similar to the e-luminexx vascular stent products that are cleared in the us.The pro code and 510k number for the e-luminexx vascular stent products is identified.Manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent delivery system was not returned and photos were not available for evaluation.Based on the information available, the reported inability of the delivery system to be delivered to the target lesion and material deformation can not be confirmed.The investigation is closed with inconclusive result for "material deformation/positioning problem".A definite root cause of the reported incident cannot be identified.Labeling review: in reviewing the relevant labeling for this product, it was found that the instructions for use sufficiently address the potential risks.With regards to general warnings, the instructions for use states that "should unusual resistance be felt at any time during the procedure, the entire system (introducer sheath or guiding catheter and stent delivery system) should be removed as a single unit".Regarding potential damages, the instructions for use states: "visually inspect the bard e-luminexx vascular stent to verify that the device has not been damaged due to shipping or improper storage.Do not use damaged equipment".Regarding preparation the instructions for use states that "prior to inserting the delivery catheter over the guidewire, the system must be flushed with sterile saline at the two female luer ports until saline drips from the distal tip of the catheter.Flushing lubricates the surface between the inner and outer catheters".Regarding procedural access, the ifu states "gain access to the treatment site utilizing appropriate accessory equipment compatible with the 6fr bard® e¿luminexx® vascular stent system.(1) via the femoral route, insert a 0.035¿ (0.89 mm) guidewire under fluoroscopic guidance through the appropriate introducer sheath or guiding catheter and pass the lesion.(2) the delivery system requires a minimum 6fr introducer sheath".The packaging pictograms indicate an introducer size of 6f and a 0.035" guidewire.(expiry date: 06/2022).
|