• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JETSTREAM XC; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JETSTREAM XC; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 45007
Device Problem Suction Problem (2170)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2021
Event Type  malfunction  
Event Description
It was reported that the aspiration rate was not normal.The target lesion was located in the left distal superficial femoral artery (sfa) to the distal popliteal.A 2.1mm jetstream xc was used to treat the distal sfa for 140 seconds.The.014x300 thruway guidewire was removed from the guard, then the jetstream was advanced to the distal popliteal for treatment.During treatment of the second lesion, the guidewire was placed back in the guard; however, the atherectomy portion of the device lost power and the device never started spinning.Aspiration was not working at the same rate as normal.The pump did not work when attempting to rex the device.The device was removed from the patient over the guidewire.Another of the same device was used to complete the procedure.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM XC
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13319960
MDR Text Key284291910
Report Number2134265-2022-00358
Device Sequence Number1
Product Code MCW
UDI-Device Identifier08714729889892
UDI-Public08714729889892
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/22/2023
Device Model Number45007
Device Catalogue Number45007
Device Lot Number0027713797
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/29/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/22/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age80 YR
Patient SexMale
Patient RaceWhite
-
-