Medtronic received information that 4 years and 6 months post implant of this 27mm aortic bioprosthetic valve, it was explanted and replaced with a 25mm aortic bioprosthetic valve. the reason for the replacement was aneurysmal dilatation.It was reported that this was a true aneurysm with the suture lines all intact and "the disruption was just above the leaflets, with huge sinus dilatation carrying the coronary buttons with them".No additional adverse patient effects were reported.
|
Product analysis: upon receipt at medtronic's quality laboratory, visual examination revealed that the valve appeared flat or collapsed and appears to have been modified with sinuses removed.All leaflets were flexible.The edges appeared rolled and smooth suggesting ¿adventitial hemorrhage¿.A hole, possibly an abrasion, was observed on the inflow belly of leaflet 2.All leaflets appeared to have been in a partially closed position and commissures were intact.Tan thrombotic appearing host tissue lined the belly of leaflet 3.A layer of off-white pannus noted on the inflow margin of attachments of all cusps.Radiography did not reveal calcification on the valve and/or host tissue.Conclusions: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.H3: device evaluated? updated h6: updated the codes medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|