• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problems Electrical /Electronic Property Problem (1198); Audible Prompt/Feedback Problem (4020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2022
Event Type  malfunction  
Manufacturer Narrative
The freedom driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.Ce (b)(4) initial.
 
Event Description
The customer, a syncardia authorized distributor, reported that the freedom driver exhibited a fault alarm while supporting a patient.There was no reported adverse patient impact.The customer also reported that the patient was subsequently changed to a backup driver.
 
Manufacturer Narrative
The driver's alarm history was reviewed and revealed multiple "19" (bottom pressure too low) alarm codes.The driver did not pass functional testing as it exhibited a continuous alarm.A "19" fault code is usually an indication of a malfunction of the u22 pressure sensor on the main printed circuit board assembly (pcba).A known functioning main pcba was installed in the driver and retested with no alarms.Additionally, the suspect main pcba was installed in a test driver which immediately exhibited an alarm when tested.The root cause of the customer-reported fault alarm was determined to be a malfunction of the u22 pressure sensor on the freedom main pcba.This issue will be tracked and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
sagar pimpalwar
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key13347291
MDR Text Key284969545
Report Number3003761017-2022-00012
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 01/14/2022
Initial Date FDA Received01/25/2022
Supplement Dates Manufacturer Received01/14/2022
Supplement Dates FDA Received08/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/21/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age49 YR
Patient SexMale
-
-