NATUS MEDICAL INCORPORATED ISOLATION TRANSFORMER - 110V ERGOJUST CART Back to Search Results
Model Number ISO-NA-EJ
Device Problems Thermal Decomposition of Device (1071); Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2022
Event Type  malfunction  
Event Description
Medical grade ups with isolation - 110v ergojust cart.The technician plugged in the power cord and smelt burning.Immediately unplugged and put out of service.No patient involved.No injuries.
 
Manufacturer Narrative
Initial report (ref complaint (b)(4)).The customer reported power plug design is bad and keeps getting burned.Customer reported found note on machine xlicumon09 s/n (b)(4)(computer) that isolation transformer sparked and smoked at plug.Customer reported no one was injured.Technical service requested customer return device to natus.Request photo of rating plate of device with manufacture date and part number.Did device have 10 foot cable or 15-foot cable? per questionnaire: date of event (b)(6) 2022.Device: ametek powervar 3.0, s/n (b)(4).Technician plugged in power cord, smelt burning, immediately unplugged and put out of service.A video eeg was going to be performed.Equipment was not on patient.Procedure did not take place.Tpi-2335 created for engineering investigation.Affected part requested to be returned for investigation.Per doc (b)(4) xltek eeg psg risk analysis, hazard i.D.Line 2.1 failure of amplifier power supply/ transformer, the severity score 11 is moderate.A risk review is not required as this complaint does not describe a new failure mode or new harm and the existing hazard severity and/or probability of occurrence has not changed.
 
Event Description
Isolation transformer - 110v ergojust cart.The technician plugged in the power cord and smelt burning.Immediately unplugged and put out of service.No patient involved.No injuries.
 
Manufacturer Narrative
Follow up report 001 ref complaint (b)(4)).(b)(6) 2022 engineering shipped medical grade ups with isolation s/n (b)(4) to powervar for evaluation.Powervar failure analysis report which was closed on the 12 april 2022 noted the following details: they disassembled and inspected both nema 5-15 connectors.The main damages to the internal elements resulted in power cord shortages are: exposed wires with melted insulation inside and outside of the connector bodies.Ground and neutral wires forcefully detached from points of termination with broken pieces of strain still inside the terminals.The nature of the damages allows to conclude that the cause of shortages was an overly excessive pull-out force applied to power cords repetitively for an extended time period.No corrective actions identified due to the nature of the damages.Closure rationale:complaint verified, being tracked as a trend.Based on evaluation of this complaint there is no safety risk of harm, individual complaint related to issue stated.Complaint will be included in trending data for further review and investigation if needed.
 
Manufacturer Narrative
Follow up report (ref complaint (b)(4)).Customer reported fourth isolation transformer same issue.No one was injured.Field service technician.On-site: customer had extra isolation transformers.Could not associate isolation transformer with particular cart.Customer took first two isolation transformers with burnt plugs and replaced with spares.Returned those two and would provide tracking numbers.Isolation transforms have ten-foot cable.Two isolation transformers with burnt plugs at site.Plugs had a burn hole where ac plug clamp squeezed cable to secure it.Field service installed four ups units on four carts with isolation transformer box.Customer could not determine which cart originally had a burnt plug.The customer provided pictures of the affected part.21 mar 2022 engineering shipped medical grade ups with isolation s/n (b)(4) to powervar for evaluation.11 apr 2022 engineering confirmed correct part for this case/related cases is (b)(4).
 
Event Description
Isolation transformer - 110v ergojust cart.The technician plugged in the power cord and smelt burning.Immediately unplugged and put out of service.No patient involved.No injuries.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOLATION TRANSFORMER - 110V ERGOJUST CART
Type of Device
ISOLATION TRANSFORMER - 110V ERGOJUST CART
Manufacturer (Section D)
NATUS MEDICAL INCORPORATED
dba excel-tech ltd.
2568 bristol circle
oakville, ontarios, l6h 5s1,
CA 
Manufacturer (Section G)
NATUS MANUFACTURING LIMITED
ida business park
gort
galway h91pd92,
EI  
Manufacturer Contact
gráinne walsh
dba excel-tech
2568 bristol circle
oakville, ontario, l6h 5s1 
CA  
MDR Report Key13356792
MDR Text Key286643875
Report Number9612330-2022-00005
Device Sequence Number1
Product Code GWQ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberISO-NA-EJ
Device Catalogue NumberISO-NA-EJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received01/14/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-