|
Model Number ACT2030 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during ablation procedure, frequency pin operation is unstable (power fluctuated up and down).They used another like device and it worked fine.There was no patient injury.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted a large bend to the needle or electrode of the device.A circulation leak test was preformed and water passed through the dev ice's tubing without issue.The tubing was crimped to create back pressure, however no leaks were observed.No nicks were seen on the tubing and cable.The device passed a hipot test.The device was able to accurately detect temperature.An ablation test was preformed on porcine kidney.The kidney was dissected and the device had ablated the area with acceptable results.It was reported that the temperature of the device was unstable or fluctuating.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: the unit was mishandled or dropped.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: verify that all components are undamaged prior to use.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during ablation procedure, the frequency pin operation is unstable (power fluctuated up and down).The needle sometimes did not work during use, the temperature in the working area was sometimes normal, and sometimes it could not go up.They used another like device and it worked fine.There was no patient injury.
|
|
Search Alerts/Recalls
|
|
|