• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET FRANCE S.A.R.L. OPTIPAC-S 60 REFOBACIN BONE CEMENT R; BONE CEMENT, ANTIBIOTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET FRANCE S.A.R.L. OPTIPAC-S 60 REFOBACIN BONE CEMENT R; BONE CEMENT, ANTIBIOTIC Back to Search Results
Catalog Number 4711500396-1
Device Problems Microbial Contamination of Device (2303); Patient Device Interaction Problem (4001)
Patient Problems Bacterial Infection (1735); Unspecified Infection (1930); Implant Pain (4561); Swelling/ Edema (4577)
Event Date 12/01/2020
Event Type  Injury  
Event Description
It was reported through a clinical study that a patient underwent a primary total knee arthroplasty on (b)(6) 2019.Subsequently, the patient was revised on (b)(6) 2020 date due to deep infection.
 
Manufacturer Narrative
This is a combination product: (b)(4).List of associated devices: psn asf cr 10mm ply l 3-11 ef, reference 42-5110-005-10, batch 63535021; psn tib stm 5 deg sz f l, reference 42-5320-075-01, batch 63560762; psn fem cr cmt ccr std sz 11 l, reference 42-5026-070-01, batch 63567263; psn all poly pat ply 38mm, reference 42-5400-000-38, batch 64100059.Report source, foreign - event occurred in (b)(6).The investigation is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.
 
Manufacturer Narrative
This is a combination product: (b)(4).List of associated devices: psn asf cr 10mm ply l 3-11 ef, reference 42-5110-005-10, batch 63535021; psn tib stm 5 deg sz f l, reference 42-5320-075-01, batch 63560762; psn fem cr cmt ccr std sz 11 l, reference 42-5026-070-01, batch 63567263; psn all poly pat ply 38mm, reference 42-5400-000-38, batch 64100059.Report source, foreign - event occurred in (b)(6).The investigation is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.
 
Event Description
It was reported through a clinical study that a patient underwent a primary total knee arthroplasty on (b)(6) 2019.Subsequently, the patient was revised on (b)(6) 2020 date due to deep infection.
 
Event Description
It was reported through a clinical study that a patient underwent a primary total knee arthroplasty on (b)(6) 2019.Subsequently, the patient was revised on (b)(6) 2020 date due to deep infection.The patient went to the emergency room due to swelling of the knee after heavy work, pain in the medial compartment was observed and it was not possible to examine the stability of the knee, after a knee puncture staphylococcus lugdunensis were detected, due to this revision in two stages was performed.All bone attached components were well placed and the insert did not have any wear.
 
Manufacturer Narrative
This is a combination product (b)(4).This follow-up report is being submitted to relay additional information.The review of the medical records by the hospital showed that the infection was due to staphylococcus lugdunensis and is a commensal bacteria on human skin.The product analysis can't be performed as the product was not returned.The device manufacturing quality record indicate that the released product met all requirements to perform as intended.The review of the sterilization certificate indicates that the products were sterilized according to the specification.A complaint extract was done regarding revision due to infection: 5 complaints (5 products), this one included, have been recorded on optipac 60 refobacin bone cement r-1, reference (b)(4), from (b)(6), 2019 to (b)(6), 2022.1 complaint (1 product), this one included, has been recorded on optipac 60 refobacin bone cement r-1, reference (b)(4), batch a747c07115.According to available data, root cause of the event was unable to be determined.However, there is no evidence that the event is related to the product.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIPAC-S 60 REFOBACIN BONE CEMENT R
Type of Device
BONE CEMENT, ANTIBIOTIC
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
Manufacturer (Section G)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR   26903
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13382068
MDR Text Key285931662
Report Number3006946279-2022-00009
Device Sequence Number1
Product Code MBB
Combination Product (y/n)N
Reporter Country CodeDA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date10/31/2019
Device Catalogue Number4711500396-1
Device Lot NumberA747C07115
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/05/2022
Initial Date FDA Received01/28/2022
Supplement Dates Manufacturer Received02/01/2022
Supplement Dates FDA Received02/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/28/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
ASSOCIATED DEVICES LISTED IN H10.
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-