Model Number 173016 |
Device Problems
Break (1069); Component or Accessory Incompatibility (2897); Human-Device Interface Problem (2949)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/03/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during total laparoscopic hysterectomy procedure, when attempting to cut the suture after closing the vaginal cuff, the needle broke in half.Both pieces were retrieved.It was also noted that the device was difficult to toggle and load.A new device was used to resolve the issue.There was no patient injury.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during total laparoscopic hysterectomy, when attempting to cut the suture after closing the vaginal cuff, the needle broke in half.One half fell into patient cavity.Both pieces were retrieved.It was also noted that the device was difficult to toggle and load.A new device was used to resolve the issue.There was no patient injury.
|
|
Manufacturer Narrative
|
H6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the device was returned with blister package and with exposed blades.The device was not seated in the blister.The right blade was bent.The beveled walls of the devices were inspected through a microscope and no witness marks from the needle tip impacting the beveled walls were found.No shearing or deformation were observed around the toggle switch.The flat pin and center rod were found to be without abnormality.It was reported that the needle broke into separate pieces, the device was difficult to load and the device was difficult to toggle.The reported issues could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: the blades of the instrument were bent.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|