Model Number SCCS1002
Device Problems Display or Visual Feedback Problem (1184); Operating System Becomes Nonfunctional (2996)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 01/06/2022
Event Type  malfunction  
Event Description
It was reported that the sensica unit had a phantom touch that persisted indefinitely which caused other user input to not fully register as the device assumed that the user was touching two separate points on the screen at the same time.The screen was technically responding to the other touches and one button was able to be pressed but not deliberately, and the resulting action could not be undone.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Manufacturer Narrative
The device was not returned.The reported issue was confirmed design related.The root cause of the reported issue is being investigated.The device did not meet specifications and was influenced by the reported failure.It was unknown if the device was in use on a patient.The device history record review was not required as the complaint was addressed by existing capa.Labeling review was not required as the complaint was addressed by existing capa.H11:section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Event Description
It was reported that the sensica unit had a phantom touch that persisted indefinitely which caused other user input to not fully register as the device assumed that the user was touching two separate points on the screen at the same time.The screen was technically responding to the other touches and one button was able to be pressed but not deliberately, and the resulting action could not be undone.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSICA UO MONITOR ICU
Type of Device
SENSICA UO MONITOR
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13390038
MDR Text Key285412145
Report Number1018233-2022-00227
Device Sequence Number1
Product Code EXS
UDI-Device Identifier00801741215773
UDI-Public00801741215773
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSCCS1002
Device Catalogue NumberSCCS1002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/31/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-