BOSTON SCIENTIFIC CORPORATION ULTRAFLEX TRACHEOBRONCHIAL; PROSTHESIS, TRACHEAL, EXPANDABLE Back to Search Results
Model Number M00576530
Device Problems Break (1069); Difficult to Remove (1528); Unraveled Material (1664); Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/07/2022
Event Type  Injury  
Event Description
It was reported to boston scientific corporation on (b)(6) 2022 that an ultraflex tracheobronchial covered distal release stent was to be implanted to treat a malignant stricture in the left main stem of the lung during a tracheobronchial stent placement procedure performed on (b)(6) 2022.The patient's anatomy was not tortuous and was not dilated prior to stent placement.During the procedure, the stent was misdeployed and the stent location could not be confirmed under live fluoroscopy due to anatomical or procedural circumstances.During removal of the delivery system, the stent was dislodged into the trachea.The stent was removed from the patient with forceps.During removal of the stent, the stent was unraveled and the stent suture was broken.The procedure was not completed due to availability of the device.There were no patient complications reported due to this event.A photo provided by the complainant of the device outside the patient showed the stent was unraveled and the stent suture was broken.
 
Manufacturer Narrative
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX TRACHEOBRONCHIAL
Type of Device
PROSTHESIS, TRACHEAL, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13396780
MDR Text Key288317642
Report Number3005099803-2022-00329
Device Sequence Number1
Product Code JCT
UDI-Device Identifier08714729842330
UDI-Public08714729842330
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141584
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/17/2023
Device Model NumberM00576530
Device Catalogue Number7653
Device Lot Number0027850499
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/07/2022
Initial Date FDA Received01/31/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/17/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient RaceBlack Or African American
-
-