Catalog Number 875205
Device Problem Overheating of Device (1437)
Patient Problems Burn(s) (1757); Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
This report summarizes 47 malfunction events where a midwest e plus handpiece overheated.In 8 events, patients had a minor burn and did not require intervention.In 28 events, no injury resulted.In 11 events, the outcome is unknown despite follow-up.
 
Manufacturer Narrative
Additional serial numbers included in this report: (b)(4).47 of 47 devices were returned for evaluation.Evaluation of 1 device found it to be within specification.Evaluation of 4 devices found excessive wear due to a lack of proper maintenance and lubrication.The devices had debris build-up.The devices did not heat up during evaluation.The devices were repaired and returned to the customers.Evaluation of 13 devices found excessive wear due to a lack of proper maintenance.The devices did not heat up during evaluation.The devices were repaired and returned to the customers.Evaluation of 3 devices found excessive wear due to a lack of proper maintenance.The devices did heat up during evaluation.The devices were repaired and returned to the customers.Evaluation of 22 devices found excessive wear due to a lack of proper maintenance.The devices had debris build-up.The devices did not heat up during evaluation.The devices were repaired and returned to the customers.Evaluation of 4 devices found excessive wear due to a lack of proper maintenance.The devices had debris build-up.The devices did heat up during evaluation.The devices were repaired and returned to the customers.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST E PLUS 1:5 HIGH SPEED CONTRA ANGLE ATTACHMENT
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM  D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM   D-64625
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494593
MDR Report Key13401397
MDR Text Key289336392
Report Number9614977-2021-00084
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
PMA/PMN Number
K972436
Number of Events Reported47
Summary Report (Y/N)Y
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number875205
Patient Sequence Number1
-
-