Model Number 744000
Device Problems Output below Specifications (3004); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Device was inspected and tested.During testing the output of pk2@ 80wthermionic voltage threshold was found lower than the limit due to faulty socket select board.The current software is v3.03.The housing has multiple minor scratches can use as is.Review of fault log showed error 400 ref 26, four times: generated if a turis electrode is attached and activated without a saline solution being present, or there is an electrode connection fault.400 ref 11, one time: footswitch yellow pedal stuck.The usual cause is that the relevant footpedal is held down during power on self-test or there is a faulty footpedal.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event Description
During an asset return inspection it was found the output of pk2@ 80wthermionic voltage threshold is lower than the limit due to faulty socket select board unit.There is no patient involvement associated on this reported event.No user injury reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 15 years since the subject device was manufactured.Based on the results of the investigation, the unit was found to have lower output due to a faulty socket select board.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYRUS, PK-SP GENERATOR
Type of Device
GYRUS, PK-SP GENERATOR
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
Manufacturer Contact
kenneth pittman
800 west park drive
westborough, MA 01581
9013785969
MDR Report Key13407124
MDR Text Key294162938
Report Number3003790304-2022-00020
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00821925009271
UDI-Public00821925009271
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031085
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number744000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-