Model Number 744000 |
Device Problems
Output below Specifications (3004); Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Device was inspected and tested.During testing the output of pk2@ 80wthermionic voltage threshold was found lower than the limit due to faulty socket select board.The current software is v3.03.The housing has multiple minor scratches can use as is.Review of fault log showed error 400 ref 26, four times: generated if a turis electrode is attached and activated without a saline solution being present, or there is an electrode connection fault.400 ref 11, one time: footswitch yellow pedal stuck.The usual cause is that the relevant footpedal is held down during power on self-test or there is a faulty footpedal.Investigation is ongoing.This report will be supplemented accordingly following investigation.
|
|
Event Description
|
During an asset return inspection it was found the output of pk2@ 80wthermionic voltage threshold is lower than the limit due to faulty socket select board unit.There is no patient involvement associated on this reported event.No user injury reported.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 15 years since the subject device was manufactured.Based on the results of the investigation, the unit was found to have lower output due to a faulty socket select board.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
|
|