• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH-AT ARTIS ZEE CEILING; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH-AT ARTIS ZEE CEILING; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 10094137
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2021
Event Type  malfunction  
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zee ceiling system.During an interventional procedure, the user reported that the dsa machine suddenly began leaking water from the tube water cooler.As a result, no x-ray release was possible.There is no report of impact to the state of health of any patient or user involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause could not be determined.The investigation was performed considering complaint description, cs reports, and system history.The customer reported that on (b)(6) 2021, the system suddenly leaked water from the tube water cooler which led to the suspension of the normal interventional operation.The issue was resolved by an independent (non-siemens) service organization.Unfortunately, in this case, a more in-depth and verifiable analysis of the problem described in the complaint was not possible as the customer refused the onsite visit from siemens service.According to experts, a problem inside the x-ray tube cooling system is conceivable.A possible general error which would require corrective action of the installed base could not be identified by the investigation.The incident described was not classified as a reportable event after detailed investigation, because neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected, since the corresponding probability according to the risk analysis is in the range of inconceivable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE CEILING
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
4843231631
MDR Report Key13412400
MDR Text Key285332001
Report Number3004977335-2022-12518
Device Sequence Number1
Product Code OWB
UDI-Device Identifier04056869010052
UDI-Public04056869010052
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K181407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10094137
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received05/17/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-