ABBOTT QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3371-40QC
Device Problems Inappropriate or Unexpected Reset (2959); Compatibility Problem (2960)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2022
Event Type  malfunction  
Event Description
It was reported that the device was programmed into magnetic resonance imaging (mri) mode prior to mri exposure.The patient underwent an mri and the device was in backup vvi mode with no backup defibrillation support available following the mri.Technical support was contacted and the device was reprogrammed to resolve the event.The patient was stable and will continue to be monitored.
 
Event Description
Additional information received indicated the device was explanted due to normal elective replacement.
 
Manufacturer Narrative
The reported event of backup operation was confirmed.Analysis of device image revealed the device went into backup mode due to a power on reset.The device was successfully restored from backup mode in the field.Functional testing was performed and the telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were observed to be normal.The backup operation could not be reproduced.Additionally, a software investigation was performed and the root cause of the power on reset could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13416463
MDR Text Key284825665
Report Number2017865-2022-02272
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2016
Device Model NumberCD3371-40QC
Device Lot Number4765780
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/29/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-