• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL CLARIA MRI CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL CLARIA MRI CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTMA1D1
Device Problems Failure to Deliver Shock/Stimulation (1133); Device Sensing Problem (2917)
Patient Problems Cardiac Arrest (1762); Tachycardia (2095)
Event Date 01/20/2022
Event Type  Injury  
Event Description
It was reported that the patient was pulseless, in ventricular tachycardia (vt), and the cardiac resynchronization therapy defibrillator (crt-d) didn't shock the patient as the device classified the rhythm as non-sustained ventricular tachycardia. a possible detection reset by the device discriminator used to distinguish between rapid and gradual onset of fast ventricular rhythms was noted and may have contributed to therapy not being delivered. cardiopulmonary resuscitation was performed and the device remains in use. no further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARIA MRI CRT-D SURESCAN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key13416470
MDR Text Key284817290
Report Number9614453-2022-00309
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00763000059552
UDI-Public00763000059552
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/28/2022
Device Model NumberDTMA1D1
Device Catalogue NumberDTMA1D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/20/2022
Initial Date FDA Received02/01/2022
Date Device Manufactured07/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
383069 LEAD (X2), 694765 LEAD
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age72 YR
Patient SexMale
Patient Weight89 KG
-
-