• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Model Number 71335554
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Bacterial Infection (1735)
Event Date 09/27/2019
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
Us legal- bilateral patient.It was reported that after the patient underwent a first revision surgery of the left hip due to trunnionosis and metallosis on (b)(6) 2019, the patient kept suffering from complications related to prosthetic joint infection.The patient had issues with postop wound healing and developed infections, requiring surgical debridement twice, as well as recurrent aspirations.On (b)(6) 2019, the patient had a second revision surgery, where the acetabular component was replaced with antibiotic spacers.
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, the ¿chronic type infection-bone destructive process¿ noted intraoperatively is consistent with the pathological findings of the bacterial infections staph epi and klebsiella pneumoniae.However, both infections are highly likely of an exogenous nature and there is no evidence that our product contributed to the infection.The infected right hip cannot be ruled out as a contributing factor to left hip infection.As well it should be noted the revision report noted that the ¿60 mm smith and nephew trial¿ was implanted.It cannot be confirmed if this was a mis-report in the document and it is referring to an implant component or if the ¿trial¿ was actually used as a stage of the revision procedure.Conclusions of using the trial as an implant cannot be concluded.The patient impact beyond the revision and expected transient post-op convalescence period cannot be determined, although it is noted the patient continued to be treated for ongoing infection post revision and had a 3rd revision.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for total hip systems revealed that periprosthetic infection has been identified as a potential complication.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the sterilization records revealed the batch was sterilized within normal parameters.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to contamination, patient reaction, and post-operative healing issue.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 54MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13416575
MDR Text Key284820764
Report Number1020279-2022-00410
Device Sequence Number1
Product Code MBL
UDI-Device Identifier03596010598257
UDI-Public03596010598257
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71335554
Device Catalogue Number71335554
Device Lot Number16FM09620
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/03/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age41 YR
Patient SexMale
-
-