A us distributor contacted zoll to report that a patient passed away at home while wearing the lifevest on (b)(6) 2021.Review of the download data, indicates the patient received one inappropriate treatments in response to cpr/motion artifact.At 13:44:03, the patient was in sinus bradycardia at 20 bpm degrading to asystole.At 16:51:28, the patient received an inappropriate treatment.The patient's rhythm at the time of the treatment event was asystole with intermittent cardiac activity with cpr/motion artifact.The patient's post-shock rhythm was asystole with intermittent cardiac activity with cpr/motion artifact and electrode lead falloff.The electrode belt was disconnected at 17:59:53.The patient passed away on (b)(6) 2021 at approximately 05:29 pm.There is no indication that the inappropriate treatment caused or contributed to the patient's passing as the patient was in a non-life-sustaining rhythm prior to the treatment.
|