The electronic log file and the replaced motor were available for investigation.The reported ventilator failure could be reconstructed by means of the information stored in the log file.During the case in question, the apollo detected a deviation between the measured and the expected position of the ventilator piston.The device reacted in accordance with its implemented safety concept as it interrupted automatic ventilation, switched to man/spont mode and generated a corresponding audible and visible ventilator fail alarm.In such case both manual ventilation and monitoring functions will remain unaffected and available.Hence, the user is able to ventilate the patient manually and monitor relevant information regarding ventilation and gas delivery.Further investigation of the returned ventilator motor revealed an excessive amount of oil on the spindle and the encoder disc of the motor assembly.It was concluded that the oil temporarily disturbed the optical detection of the motor position.The replacement of the motor assembly on-site has already fixed the problem.Within the last three years, no similar cases have been reported to draeger.
|