• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA APOLLO; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA APOLLO; ANESTHESIA UNITS Back to Search Results
Model Number 8606500
Device Problems Gas Output Problem (1266); Failure to Analyze Signal (1539); Failure to Deliver (2338)
Patient Problem Insufficient Information (4580)
Event Date 01/13/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Event Description
It was reported that the unit alarmed for a ventilator failure during a case.There was no patient injury reported.
 
Manufacturer Narrative
The electronic log file and the replaced motor were available for investigation.The reported ventilator failure could be reconstructed by means of the information stored in the log file.During the case in question, the apollo detected a deviation between the measured and the expected position of the ventilator piston.The device reacted in accordance with its implemented safety concept as it interrupted automatic ventilation, switched to man/spont mode and generated a corresponding audible and visible ventilator fail alarm.In such case both manual ventilation and monitoring functions will remain unaffected and available.Hence, the user is able to ventilate the patient manually and monitor relevant information regarding ventilation and gas delivery.Further investigation of the returned ventilator motor revealed an excessive amount of oil on the spindle and the encoder disc of the motor assembly.It was concluded that the oil temporarily disturbed the optical detection of the motor position.The replacement of the motor assembly on-site has already fixed the problem.Within the last three years, no similar cases have been reported to draeger.
 
Event Description
It was reported that the unit alarmed for a ventilator failure during a case.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APOLLO
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
MDR Report Key13437839
MDR Text Key287331198
Report Number9611500-2022-00030
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier04048675229674
UDI-Public(01)04048675229674(11)170720(17)171218(93)8606500-64
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8606500
Device Catalogue Number8606500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/31/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-