Model Number 71992-01
Device Problems Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
Patient Problems Hypoglycemia (1912); Loss of consciousness (2418); Cramp(s) /Muscle Spasm(s) (4521)
Event Date 01/15/2022
Event Type  Injury  
Manufacturer Narrative
At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.Dhrs (device history record) for the libre sensor and libre sensor kits were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported receiving a "replace sensor" message upon scanning the freestyle libre 2 sensor.As a result, customer was unable to monitor glucose via sensor scan and experienced symptoms described as "back and neck pain, hard breathing, whole-body contraction", and loss of consciousness.The customer was treated with glucagon injection by his wife.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key13440979
MDR Text Key284974234
Report Number2954323-2022-03840
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date03/31/2023
Device Model Number71992-01
Was Device Available for Evaluation? No
Date Manufacturer Received01/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
Patient Weight77 KG
-
-