• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G447
Device Problems Failure to Capture (1081); High Capture Threshold (3266)
Patient Problem Cardiac Perforation (2513)
Event Date 12/17/2021
Event Type  Injury  
Event Description
It was reported that this system exhibited high pacing thresholds and loss of capture on the atrial channel.A perforation was initially suspected and imaging was performed; however, this could not be confirmed via imaging.The physician determined that the patient did not respond well to pacing in the ventricle and most likely would not benefit from cardiac resynchronization therapy (crt).The system was explanted and replaced.No additional adverse patient effects were reported.
 
Manufacturer Narrative
This product is being evaluated in our post market quality assurance laboratory.This report will be updated when evaluation is complete.
 
Event Description
It was reported that this system exhibited high pacing thresholds and loss of capture on the atrial channel.A perforation was initially suspected and imaging was performed; however, this could not be confirmed via imaging.The physician determined that the patient did not respond well to pacing in the ventricle and most likely would not benefit from cardiac resynchronization therapy (crt).The system was explanted and replaced.No additional adverse patient effects were reported.
 
Manufacturer Narrative
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13445428
MDR Text Key285036546
Report Number2124215-2022-02553
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589539
UDI-Public00802526589539
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/27/2022
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number248159
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/17/2021
Initial Date FDA Received02/03/2022
Supplement Dates Manufacturer Received02/24/2022
Supplement Dates FDA Received03/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention; Hospitalization;
Patient Age57 YR
Patient SexMale
-
-