Model Number 71362279
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event Description
It was reported that, the reamer handle fell out.As this was noticed upon field inspection, there was not patient involvement.
 
Event Description
It was reported that, the spring and lock of the reamer handle fell out.As this was noticed upon field inspection, there was not patient involvement.
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirms the spring and lock ring are detached from the device, rendering the device inoperable.The spring was not returned but the lock ring was.The device shows signs of extreme wear and use.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device was missing a component and it could not fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Misassembling after reprocessing or rough handling are potential factors that may have contributed to the reported event.Additionally we recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAMER HANDLE
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13446326
MDR Text Key285168544
Report Number1020279-2022-00471
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010456519
UDI-Public03596010456519
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71362279
Device Catalogue Number71362279
Device Lot Number09BM20002
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/17/2022
Initial Date FDA Received02/03/2022
Supplement Dates Manufacturer Received01/17/2022
03/02/2022
11/30/2022
Supplement Dates FDA Received02/04/2022
03/02/2022
12/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/26/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-