SMITH & NEPHEW, INC. GII MIS DCF ALIGN GDE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71441144 |
Device Problem
Failure to Align (2522)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/17/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference case-(b)(4).
|
|
Event Description
|
It was reported that the gii mis dcf align gde no long slides into place.As this was noticed upon field inspection, no patient was involved.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that the gii mis dcf align gde no long slides into place.As this was noticed upon field inspection, no patient was involved.
|
|
Manufacturer Narrative
|
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirms the locking mechanism is seized in place.The device shows normal signs of wear/usage.A review of complaint history revealed similar events for the listed device, but no similar events for the batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.
|
|
Search Alerts/Recalls
|
|
|