• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VDW GMBH WST 6:1 F. VDW.GOLD/SILVER; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VDW GMBH WST 6:1 F. VDW.GOLD/SILVER; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL Back to Search Results
Catalog Number V041079000000
Device Problems Device Slipped (1584); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it has been reported that vdw gold/silver won't hold file.File fell out during treatment.No injury occurred.Device will be returned for investigation.
 
Manufacturer Narrative
The device was evaluated and found to be within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WST 6:1 F. VDW.GOLD/SILVER
Type of Device
HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL
Manufacturer (Section D)
VDW GMBH
bayerwaldstrasse 15
munich bayern 81737
GM  81737
Manufacturer (Section G)
VDW GMBH
bayerwaldstrasse 15
munich bayern 81737
GM   81737
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key13453137
MDR Text Key287580575
Report Number9611053-2022-00472
Device Sequence Number1
Product Code EGS
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberV041079000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date02/02/2022
Initial Date Manufacturer Received 02/02/2022
Initial Date FDA Received02/04/2022
Supplement Dates Manufacturer Received02/02/2022
Supplement Dates FDA Received03/31/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-