Model Number CDHFA500Q |
Device Problem
Component Missing (2306)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/19/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the patient presented for a generator change procedure.During the procedure, it was noted that the left ventricular lead could not be screwed into the left ventricular port of the implantable cardioverter defibrillator because the set screw appeared to be missing.The device was removed and replaced.The patient was in stable condition.
|
|
Event Description
|
It was reported that the patient presented for a generator change procedure.During the procedure, it was noted that the left ventricular lead could not be screwed into the left ventricular port of the implantable cardioverter defibrillator because the set screw appeared to be missing.The device was removed and replaced.The patient was in stable condition.
|
|
Manufacturer Narrative
|
The reported event of a missing set screw could not be confirmed.No set screws were missing from the header.Visual inspection of the header found the left ventricular (lv) set screw to be dislodged from the set screw bore.The dislodged set screw is due to over loosening the set screw.Test leads were able to be inserted, secured, and loosened normally in all chambers using the received set screws.
|
|
Search Alerts/Recalls
|