• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problems Failure to Cut (2587); Patient Device Interaction Problem (4001)
Patient Problems Abrasion (1689); Scar Tissue (2060)
Event Date 01/31/2022
Event Type  Injury  
Manufacturer Narrative
This complaint has been recorded under (b)(4).Once investigation is completed a supplemental/final report will be submitted.
 
Event Description
It was reported that during surgery the device did not produce a smooth harvest as the device produced skip lesions and did not follow selected depth.The entire graft was shredded and an additional graft was harvested from the patient.An additional device was used to complete procedure and as a result a 5-10 minute delay occurred where the patient was under anesthesia.
 
Manufacturer Narrative
This complaint has been recorded under (b)(4).Review of the most recent repair record determined the unit was found to be out of calibration.The thickness control lever and vespel and semi circle bearings were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key13472330
MDR Text Key285268007
Report Number0001526350-2022-00115
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024375901
UDI-Public(01)00889024375901(11)210928(10)652390667
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number65239067
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/28/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-