• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC AUTOMATRIX SHAFT; INSTRUMENTS, DENTAL HAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC AUTOMATRIX SHAFT; INSTRUMENTS, DENTAL HAND Back to Search Results
Catalog Number 62422603
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
In this event it has been reported that automatrix shaft broke during treatment.No injury occurred.
 
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Manufacturer Narrative
Capa-2021-375 opened to address weld failures for product manufactured by sarasota since december 2020 (date code 1220).Complaint is considered substantiated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOMATRIX SHAFT
Type of Device
INSTRUMENTS, DENTAL HAND
Manufacturer (Section D)
DENTSPLY LLC
38 west clarke avenue
milford DE 19963
Manufacturer (Section G)
DENTSPLY LLC
38 west clarke avenue
milford DE 19963
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key13479880
MDR Text Key287902121
Report Number2515379-2022-00038
Device Sequence Number1
Product Code DZN
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number62422603
Device Lot Number04827428
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2022
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date02/04/2022
Initial Date Manufacturer Received 02/04/2022
Initial Date FDA Received02/08/2022
Supplement Dates Manufacturer Received02/04/2022
Supplement Dates FDA Received06/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-