VERATHON MEDICAL ULC GLIDESCOPE AVL VIDEO BATON 3-4; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0570-0307
Device Problems Erratic or Intermittent Display (1182); Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2022
Event Type  malfunction  
Manufacturer Narrative
The customer declined the option to have their glidescope avl video baton 3-4 returned to verathon for evaluation.Upon review of the device history for the serial number (b)(4), it was determined that the device was manufactured on january 27, 2015 and is past the two (2) year expected product life as outlined in the glidescope operations and maintenance manual (omm).Since the device was not returned to verathon for evaluation, the cause could not be determined, but it is likely that the age of the device, seven (7) years, caused or contributed to the event.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event Description
The customer reported that during a patient procedure, the glidescope avl video baton 3-4 had a loose tip and intermittent picture.The procedure was completed using a different glidescope avl video baton 3-4 and glidescope avl monitor.No delay in the procedure or harm to the patient or user was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE AVL VIDEO BATON 3-4
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key13481420
MDR Text Key287187607
Report Number9615393-2022-00025
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0570-0307
Device Catalogue Number0570-0313
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/28/2022
Initial Date FDA Received02/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-