Model Number VPR-GW-FLEX14
Device Problem Material Separation (1562)
Patient Problems Foreign Body In Patient (2687); Foreign Body Embolism (4439)
Event Date 01/13/2022
Event Type  Injury  
Manufacturer Narrative
Analysis of the reported device is in progress.A supplemental report will be submitted when the device analysis is completed.Csi id: (b)(4).
 
Event Description
Orbital atherectomy treatment was performed for one treatment pass on each speed.During the removal process, it was observed that the tip of the viperwire guide wire had fractured off and embolized the peroneal artery.The fragment was retrieved with a snare device and balloon angioplasty restored flow.The patient was reported to be doing well.
 
Manufacturer Narrative
The reported guide wire was returned to csi for analysis.A visual examination confirmed that the guide wire was fractured, and the distal fragment was returned in a non-csi catheter.Scanning electron microscopy analysis showed evidence of fatigue in the spring tip support coil.At the conclusion of the device analysis investigation, the report that the guide wire had fractured was confirmed, however the exact root cause of the fracture could not be determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Type of Device
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul MN 55112
Manufacturer Contact
brittany leider
1225 old highway 8 nw
saint paul, MN 55112
6512591600
MDR Report Key13498292
MDR Text Key285345535
Report Number3004742232-2022-00033
Device Sequence Number1
Product Code MCW
UDI-Device Identifier10852528005718
UDI-Public(01)10852528005718(17)230531(10)380877-1
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190634
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberVPR-GW-FLEX14
Device Catalogue Number7-10041-03
Device Lot Number380877-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/13/2022
Initial Date FDA Received02/09/2022
Supplement Dates Manufacturer Received02/11/2022
Supplement Dates FDA Received02/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
Patient SexMale
Patient Weight40 KG
Patient RaceWhite
-
-