The reported guide wire was returned to csi for analysis.A visual examination confirmed that the guide wire was fractured, and the distal fragment was returned in a non-csi catheter.Scanning electron microscopy analysis showed evidence of fatigue in the spring tip support coil.At the conclusion of the device analysis investigation, the report that the guide wire had fractured was confirmed, however the exact root cause of the fracture could not be determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
|