Catalog Number CON-HL-390
Device Problem Power Problem (3010)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that there were power switch issues.
 
Manufacturer Narrative
Other text: additional information:d4 udi is unknown.No product information has been provided to date.D5, d10, h3 and h6.This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.Cracked tank cover, wear and tear damaged enclosure and front cover.Faded line cord and outdated printed circuit board (pcb) and power switch.The technician plugged line cord into power source and turned on the power switch and there was no power.The front cover and pcb were removed to check the connection between the power switch and pcb.The reported issue was confirmed.The device would not power on.The root cause of the reported issue was found to the power switch and pcb was damage caused by wear and tear by the customer and poor design.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.A corrective and preventative action has been opened to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key13518517
MDR Text Key285532892
Report Number3012307300-2022-03278
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/09/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-