SYNTHES GMBH 2.4MM TI VA-LCP VOLAR RIM DSTL RAD PL/6H HD/5H SHFT/RT-STER; PLATE, FIXATION, BONE
|
Back to Search Results |
|
Catalog Number 04.115.750S |
Device Problems
Break (1069); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/14/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional product code: hwc complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes (b)(4) reports an event as follows: it was reported that on (b)(6), 2021, the patient underwent an open reduction internal fixation (orif) for fracture of distal radius.The surgeon bent the second row of the plate on the distal side and the plate broke immediately without resistance.Another plate was used in the procedure, and the surgery was completed within a thirty (30) minute delay.No fragments were left in the patient.The patient condition was reported as stable.This report is for a distal radius plate.This is report 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: d4; d9; h4 h3, h6: the subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: product code: 04.115.750s; lot number: 353p844; manufacturing site: mezzovico; release to warehouse date: september 13, 2021; expiration date: september 1, 2031.A manufacturing record evaluation was performed for the sterile lot number, and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that va-lcp volar rim dist-rad-pl2.4 r shaft has a broken and bent conditions, these conditions are related with the customer complaint.No dimensional inspection was performed due to post-manufacturing damage and device geometry.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the broken and bent conditions observed on the a-lcp volar rim dist-rad-pl2.4 r shaft would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|