Model Number V60
Device Problem Increase in Pressure (1491)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/28/2022
Event Type  malfunction  
Event Description
It was reported to philips by a customer that the unit had a high-pressure regulation error.Based upon the information provided, the device was in clinical use providing therapy at the time of the event reported.No harm or injury has been indicated or alleged.The device was evaluated remotely by a philips remote service engineer (rse).Upon further inspection and review of the device, the biomed stated the respirator therapist reported that while trying to overcome excessive leak due to heavy beard the unit alarmed for high pressure and would not clear.Error codes 1205 (pressure regulation high) and 1206 (high inspiratory pressure) logged.The rse advised the biomed to perform full pvt and address any issues before placing into service.It is unknown if any parts or repairs have been conducted about the alleged complaint.Attempts to obtain further information are currently pending.
 
Manufacturer Narrative
Attempts have been made to acquire additional information, but no further information has been made available.If pertinent information is later received, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key13524601
MDR Text Key287018171
Report Number2031642-2022-00440
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/28/2022
Initial Date FDA Received02/14/2022
Supplement Dates Manufacturer Received03/28/2022
Supplement Dates FDA Received04/21/2022
Date Device Manufactured05/09/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-