Model Number M00535900
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2022
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that an ultratome xl was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, "the cutting wire was fractured." it was reported that the cutting wire was broken.No part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Manufacturer Narrative
Block h6 (device codes): medical device problem code a0401 captures the reportable event of cutting wire broken.Block h10: the returned ultratome xl was analyzed, and a visual evaluation noted that there were no visual problems with the device.The device was observed under magnification and the cutting wire was in good condition.A functional evaluation was performed by bowing the device outside and inside the scope.The device bowed and returned from bowed position as intended.No other problems with the device were noted.The reported event of cutting wire break was not confirmed.Upon analysis, it was determined that the device did not have any visual damage, also under magnification it was possible to observe that the cutting wire was in good condition.During functional inspection, the cutting wire bowed as intended, indicating that the cutting wire was not broken.The returned unit did not show evidence of either the alleged problem(s) or any defect which could have contributed to the event.Based on all gathered information, the most probable root cause of this complaint is no problem detected.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.Block h11 (correction): block b5 (describe event or problem) has been corrected, indicating the anatomy location involved during the procedure.
 
Event Description
It was reported to boston scientific corporation that an ultratome xl was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, "the cutting wire was fractured." it was reported that the cutting wire was broken.No part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATOME XL
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13535909
MDR Text Key288374970
Report Number3005099803-2022-00705
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729103257
UDI-Public08714729103257
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K930022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/23/2023
Device Model NumberM00535900
Device Catalogue Number3590
Device Lot Number0027197938
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/24/2022
Initial Date FDA Received02/15/2022
Supplement Dates Manufacturer Received03/02/2022
Supplement Dates FDA Received03/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/23/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52 YR
Patient SexFemale
Patient Weight62 KG
-
-