BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, STER; SYSTEM/DEVICE, PHARMACY COMPOUNDING
|
Back to Search Results |
|
Catalog Number H93811 |
Device Problems
Fluid/Blood Leak (1250); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/24/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the tubing of three (3) sterile repeater pump tube sets leaked.The tube sets seemed to be a bit longer in length in the section that sat in the repeater pump housing, causing a leak.This was identified during setup.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the tubing of three (3) sterile repeater pump tube sets leaked.The tube sets seemed to be a bit longer in length in the section that sat in the repeater pump housing, causing a leak.This was identified during setup.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information was added to h3, h4, h6 and h10.H4: the lot was manufactured from june 24, 2021 ¿ june 25, 2021.H10: the actual devices were not available; however, two (2) companion samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed in both samples and no leak was observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|