Model Number N/A |
Device Problems
Tear, Rip or Hole in Device Packaging (2385); Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/07/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that a stem was found to have punctured the packaging and a black stain was seen on the package.A new device was opened to complete the surgery.There was no reported harm or injury to a patient.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
(b)(4).Report source (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for the investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Evaluation of the returned product confirmed there is white debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging and debris consistent with the appearance of porous coating inside the sterile barrier.There reported damage to the sterile pouch has not compromised the sterility of the product.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|